UV/Visible Double Beam Spectrophotometer / ´õºí ºö ºÐ±¤±¤µµ°è |
|
Optional Accessory |
|
|
|
|
|
|
|
Cat. No. |
Model |
Description |
S23-07-150 |
6850 |
UV/Visible Double BeamSpectrohotometer |
Cat. No. |
S23-07-150 |
Model |
6850 |
Wavelength Range
| 190 to 1100 nm |
Wavelength Resolution
| 0.1 nm |
Wavelength Accuracy
| ¡¾0.3 nm (at 0.5 and 1 nm bandwidth) ¡¾0.5 nm (at 2, 4 and 5 nm bandwidth) |
Wavelength Reproducibility
| ¡¾ 0.2 nm |
Spectral Bandwidth
| Variable 0.5, 1, 2, 4, 5nm |
Photometric Range
| - 0.3 to 3.0A 0 to 200%T |
Photometric Accuracy
| ¡¾0.002A (0- 0.5A) ¡¾0.3% T (0 - 100%T) |
Photometric Reproducibility
| ¡¾0.001 Abs (0 to 0.5 Abs) ¡¾0.002 Abs (0.5 to 1.0 Abs) 0.15%T (0 - 100%T) |
Resolution
| 0.1%T, 0.001A |
Stray Light
| < 0.05%T at 220 and 360nm |
Noise
| 0.0005A |
Stability
| ¡¾0.001A/h at 500nm after 15 min warm up |
Multi-Wavelength
| Up to 10 wavelengths, up to 20 wavelengths with PC software |
Calculations
| Ratio, difference, formulae with factors |
Spectrum Range
| Any range between 190 and 1100nm |
Scan Speed
| 100 to 2000nm/min |
Scan Interval
| 0.1, 0.2, 0.5, 1, 2 or 5nm |
Analysis
| Auto peaks and valleys, zoom. Addition, subtraction, peak ratios, smoothing,
area under curve, wavelength table, derivatives, overlay with PC software |
Kinetics
| Up to 12 hours with time intervals of 0.1, 0.2, 0.5, 1, 2, 5, 10 or 30 seconds |
Analysis
| Slope and formula of line of best fit between any two points |
Quantitation Points
| Up to 3 wavelengths |
Quantitation Calibration
| Blank with up to 10 standards or factor |
Concentration Range
| 0 - 99999 |
Calibration
| Blank with standards or factor |
DNA/ RNA and Protein
| DNA/RNA Ratio, concentration, A320 correction |
Light Source
| Tungsten halogen and Deuterium lamps |
Lamp Changeover
| 325 to 370nm selectable |
Outputs
| USB and parallel |
Operating System:
| Windows 2000, XP, Vista, Windows 7 |
Electrical Supply
| 120VA, 220/110V, 50/60Hz |
Size (W x D x H)
| 600¡¿450¡¿200 mm |
Weight |
22 kg |
Optional Accessory
Cat. No |
Model |
Description |
S23-07-501 |
685131 |
Water Heated 10¡¿10 mm Single Cell Holder |
S23-07-505 |
685005 |
10 to 100 mm Path Length Cuvette Holder |
S23-07-509 |
685304 |
Micro Cuvette Holder |
S23-07-513 |
685401 |
8 Position Automatic Cell Holder |
|
|
|
|